Pt100 sensorFor steam chambers (autoclaves)
Design
Pt100 detector mounted in protection tube with cannula-like (injection needle) probe tip. The transition to cable is sealed with silicon and silicon hose. All designs with Φ 3.0 mm probe diameter.
Example of use
Temperature measurements of goods and atmosphere in steam chambers (steam autoclaves).
Maximum temperature
Probe: 180 °C
Cable: 180 °C
Sensing element
Pt100 detector, IEC60751 class A. Single or dual.
| Part number | Weight (kg) | Probe design | Measuring circuits | Signal cable type Lead area (mm2) | Connector |
| PWZP2T-1942 | 0.5 | Angled | Double | TEXSITW 8x0.25 | Stripped wires |
| PWZP2T-1941 | 0.5 | Straight | Double | TEXSITW 8x0.25 | Stripped wires |
| PWZPT-1942 | 0.3 | Angled | Single | TEXSITW 4x0.25 | 4-pole AMP |
| PWZPT-1941 | 0.3 | Straight | Single | TEXSITW 4x0.25 | 4-pole AMP |